B. The copper concentration in a contaminated salt sample was determined by flame atomic absorption spectrometry. In the analysis, a 10.00-g sample containing an analyte is transferred to a 250-mL volumetric flask and diluted to volume. A series of 10.0 mL of the diluted sample solution was pipetted into each of five 50.0-mL volumetric flasks. Various volumes of a standard containing 12.2 ppm Cu were added to the flasks, and the solutions were then diluted to volume. The corresponding absorbances of each of the solutions are given below: Unknown, mL 10.0 10.0 10.0 10.0 10.0 Standard, mL 0.0 10.0 20.0 30.0 40.0 Absorbance 0.201 0.292 0.378 0.467 0.554 What type of standardization (calibration) technique is used in this problem? What do you think is the main reason why this standardization technique is used?

Appl Of Ms Excel In Analytical Chemistry
2nd Edition
ISBN:9781285686691
Author:Crouch
Publisher:Crouch
Chapter12: Spectrochemical Methods
Section: Chapter Questions
Problem 12P
icon
Related questions
Question
B. The copper concentration in a contaminated salt sample was determined by flame atomic absorption
spectrometry. In the analysis, a 10.00-g sample containing an analyte is transferred to a 250-mL
volumetric flask and diluted to volume.
A series of 10.0 mL of the diluted sample solution was pipetted into each of five 50.0-mL volumetric
flasks. Various volumes of a standard containing 12.2 ppm Cu were added to the flasks, and the
solutions were then diluted to volume. The corresponding absorbances of each of the solutions are
given below:
Unknown, mL
10.0
10.0
10.0
10.0
10.0
Standard, mL
0.0
10.0
20.0
30.0
40.0
Absorbance
0.201
0.292
0.378
0.467
0.554
What type of standardization (calibration) technique is used in this problem? What do you think is the
main reason why this standardization technique is used?
Transcribed Image Text:B. The copper concentration in a contaminated salt sample was determined by flame atomic absorption spectrometry. In the analysis, a 10.00-g sample containing an analyte is transferred to a 250-mL volumetric flask and diluted to volume. A series of 10.0 mL of the diluted sample solution was pipetted into each of five 50.0-mL volumetric flasks. Various volumes of a standard containing 12.2 ppm Cu were added to the flasks, and the solutions were then diluted to volume. The corresponding absorbances of each of the solutions are given below: Unknown, mL 10.0 10.0 10.0 10.0 10.0 Standard, mL 0.0 10.0 20.0 30.0 40.0 Absorbance 0.201 0.292 0.378 0.467 0.554 What type of standardization (calibration) technique is used in this problem? What do you think is the main reason why this standardization technique is used?
Expert Solution
trending now

Trending now

This is a popular solution!

steps

Step by step

Solved in 2 steps

Blurred answer
Knowledge Booster
Spectroanalytical Methods
Learn more about
Need a deep-dive on the concept behind this application? Look no further. Learn more about this topic, chemistry and related others by exploring similar questions and additional content below.
Similar questions
  • SEE MORE QUESTIONS
Recommended textbooks for you
Appl Of Ms Excel In Analytical Chemistry
Appl Of Ms Excel In Analytical Chemistry
Chemistry
ISBN:
9781285686691
Author:
Crouch
Publisher:
Cengage
Principles of Instrumental Analysis
Principles of Instrumental Analysis
Chemistry
ISBN:
9781305577213
Author:
Douglas A. Skoog, F. James Holler, Stanley R. Crouch
Publisher:
Cengage Learning
Fundamentals Of Analytical Chemistry
Fundamentals Of Analytical Chemistry
Chemistry
ISBN:
9781285640686
Author:
Skoog
Publisher:
Cengage